12.
APPENDIX
A.
Specification Modbus via RS485 specification
General information
•
The Modbus (slave function) requires the firmware version V1.01.70 dated 17.11.2014 or
later.
•
The analysers are able to work as modbus slave using the RS232 or RS485 port (possibly
with external RS232/RS485 adapter)
•
supports RS485 interface with 2/4 wires (half/full duplex)
•
supports only the binary Modbus protocol (RTU)
•
supports modbus command
Read Holding Register
(command no 3)
•
supports modbus command
Read Input Register
(command no 4)
•
the slave modbus address is user definable from 1 to 238
•
communication parameter are user definable as follows:
- 9600 baud
- 19200 baud
- even parity and 1 stop bit
- no parity and 2 stop bits
•
Multi byte values are transmitted in Motorola
®
byte order (Big-Endian). Only the CRC16 at
the end of each frame is transmitted in Intel
®
byte order (Little-Endian).
In case you need Little-Endian byte order in the master's system:
- 16bit values (occurs only in the frame): swap bytes 0<=>1
- 32bit values (occurs only in the data): swap bytes 0<=>3 and swap bytes 1<=>2
•
All adresses written in this document are decimal (not hexa-decimal)
•
All readable data are 32bit values, therefore the analyzer only accepts even addresses end
even number of registers to be read.
•
The maximal number of 32bit-values to be read with one single read command is 63
(126 modbus registers)
•
Data types (used in table below):
U32
32 bit unsigned integer value (0…4.294.967.
295)
FL
32 bit floating point value (reads -1E38, when not available).
Содержание SWG100 CEM
Страница 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Страница 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Страница 98: ...MRU GmbH SWG 100 User Manual Page 98 of 155 Hardware state tests 1 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 1 ...
Страница 102: ......
Страница 135: ...MRU GmbH SWG 100 User Manual Page 135 of 155 13 3 Option O2 paramagnetic cell measurement 0 25 00 ...
Страница 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Страница 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Страница 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...