7.0 MAINTENANCE AND TROUBLE SHOOTING
7.1
General
The DP Series power supplies consist of a multistage power processing system. Because of its
complexity, it is highly recommended that all repairs be performed by the factory or qualified
power supply technician. Before attempting maintenance or repair, the technician should be
familiar with the components of the systems and the theory of operation. Some basic test
equipment is also necessary: source of ac power, means of loading the power supply, dc voltmeter
with accuracy and resolution better than the unit specifications, and an oscilloscope. The chart in
Section 7.2 should aid in finding operational problems.
Caution: When servicing the power supply, dangerous voltage levels exist.
All ac and dc capacitors should be discharged. Be especially careful of
person and equipment when measuring primary circuitry since this is at line
potential.
7.2
Trouble Shooting Guide
1.
Fuse F1, F2, or F3 blows when the power supply is turned on. Power supply has a short on
the primary side.
a.
Check diode bridge DB1.
b.
Check for failed power semiconductor on the Chopper Module.
c.
Check for isolation to ground on the input side of the power supply.
2.
Over current trips and power supply cannot be reset. An auxiliary over current detector
limits input dc link current. Exceeding safe levels will cause an over current diagnostic
condition that cannot be reset.
a.
Restart the supply. Transient voltages on the input voltage could have caused the
problem.
b.
Check diodes D1 through D4 on the secondary side of transformer T1.
c.
Check for a shorted winding on transformer T1.
d.
Check for failed IGBT’s in the Chopper and Inverter Modules.
3.
Unit goes to high output immediately after starting.
a.
Check for loose connectors internal to the power supply.
4.
High ripple voltage.
a.
Check the mains voltage for balance and magnitude.
110
Содержание DP SERIES II
Страница 1: ...OPERATING AND SERVICE MANUAL DP SERIES II DC POWER SUPPLIES...
Страница 2: ......
Страница 3: ...MAGNA POWER ELECTRONICS INC 39 ROYAL ROAD FLEMINGTON NJ 08822 January 6 2010...
Страница 4: ......
Страница 6: ...improper use of the equipment and accessories ii...
Страница 28: ...Figure 1 2 DP Series air cooled 10 kW to 30 kW package drawing with D Version front panel 15...
Страница 29: ...Figure 1 3 DP Series water cooled 10 kW to 30 kW package drawing with D Version front panels 16...
Страница 39: ...Figure 2 3 DPC Series controls and indicators Energizes control circuits without turning on main power 26...
Страница 44: ...Figure 3 2 Set voltage output at memory displayed Figure 3 3 Set current output at memory displayed 31...
Страница 45: ...Figure 3 4 Set over voltage trip at memory displayed Figure 3 5 Set over current trip at memory displayed 32...
Страница 47: ...Figure 3 7 Set remote sense Figure 3 8 Set internal control 34...
Страница 48: ...Figure 3 9 Set external control Figure 3 10 Set rotary input 35...
Страница 49: ...Figure 3 11 Set keypad input Figure 3 12 Set external program input 36...
Страница 50: ...Figure 3 13 Set remote input Figure 3 14 Set external interlock 37...
Страница 53: ...Figure 3 15 Calibration 40...
Страница 56: ...Figure 3 17 Set remote sense Figure 3 18 Set internal control Figure 3 19 Set external control 43...
Страница 57: ...Figure 3 20 Set rotary input Figure 3 21 Set external program input Figure 3 22 Set remote input 44...
Страница 61: ...Figure 3 24 Calibration 48...
Страница 63: ...Figure 3 25 a Local and b remote sensing 50...
Страница 64: ...Figure 3 26 External a resistive b voltage and current programming 51...
Страница 111: ...98...
Страница 113: ...Figure 5 1 Configuration setup Figure 5 2 GPIB communications setup 100...
Страница 118: ...Figure 5 7 Register Panel Figure 5 8 Calibration Panel 105...
Страница 119: ...Figure 5 9 Firmware Panel Figure 5 10 Modulation Panel for Version D models only 106...
Страница 121: ...Figure 6 1 Block diagram 108...
Страница 128: ...source Figure 8 2 Leadless remote sensing Figure 8 3 IV characteristics for a typical photovoltaic array 115...
Страница 139: ...Figure B 1 Information Panel Figure B 2 Configure Panel 126...
Страница 140: ...Figure B 3 Reboot in Progress Panel Figure B 4 Web Control Panel 127...
Страница 145: ...Figure D 1 RS485 a 2 wire and b 4 wire connections 132...