(1)Sterilization for:
Air Turbine Hand pieces
Micromotor Attachments
Micromotor Motor Cover
Scaler Tips and Hand pieces
This depends on the model; refer to the separate for
information about the instruments installed on your unit.
If one of the instruments in the dentist’s holder has
been disconnected, the other instruments can still
be used by releasing the Hand piece Lock.
(See [5] 3. (7) 1) in this manual.)
(2)Sterilization for:
Luna Vue ED Light: Handle Covers
Vacuum Syringe: Tips and Body
Three-way Syringe: Nozzle and Body
Saliva Ejector: SE Nozzle and Body
Dentist’s Tray: Silicone Handle Cover
1)Detachment o f Autoclavable Components
Refer to section [5] Operation in this manual.
2)Autoclave Temperatures
135
℃
(maximum). Please refer to belowing recommendation.
[6]Maintenance 1.
ADVICE
(3)Silicone Tray Sheet and Medicinal Bottle Holder
1)Autoclave Temperatures
Recommence condition:134
℃
, for 5min. (for the gravity dehydration type autoclave)
(for the gravity dehydration type autoclave)
Autoclave Parts
Temperature
Time
Vacuum Syringe VS90 Body/Chip
134
℃
5mins
Vacuum Syringe VS14 Body/Chip
121
℃
60mins
Saliva Ejector
SE14S Body/SE
Nozzle (disassembled)
134
℃
5mins
Saliva Ejector SE90S Body/SE Nozzle (disassembled)
134
℃
10mins
Threeway Syringe WS97 Body/Nozzle (disassembled)
134
℃
10mins
Threeway Syringe WS201 Nozzle (disassembled)
134
℃
10mins
Luna Vue ED Light/Dentist’s Tray: Handle Cover
134
℃
5mins
<Recommendation condition of Autoclave Parts>