55
(6)Vacuum Syringe VS-90QT
<Detachable body>
1)Operation
This syringe begins to operate when it is taken off its holder on the
assistant’s side and stops when it is returned to its holder.
(The timing for stopping can be set; see 2) below)
When the syringe has been pulled out, it can be turned off and on with
the vacuum switch.(See[5]4.(7)of this manual)
The syringe will not operate if it is pulled out while
the auto cup filler is operating.
2)Settings
Settings are shown below:
>>Display Settings
(See[5]3.(7) 7) of this manual.)
(A)Timing to stop suction (immediate or delayed).
3)L-shaped Tip
*
a)Put the rubber tip onto the L-shaped Tip.
b)Line up the bump on the tip with the groove inside the body and slide
it in.
4)Syringe Body Removal
The end of the syringe body is autoclave and can be removed.
Hold the main part of the syringe at the bottom and unscrew the body.
To adjust the angle of the syringe, loosen the tube connector and pull
it down. Then adjust the angle (the mark on the connector shows the
left position).
Raise the connector back up and retighten it.
To reattach the body, screw it back on as far as it will go.
Rotate the body to take it off. Do not twist the tube.
>>Otherwise, the instrument could be damaged.
Syringe Body
Syringe Tube
Connector
Mark
Mark
CAUTION
ADVICE
[5]Operation 6.(6)
Vacuum Switch
Rubber Tip L-shaped Tip
Groove
Bump