85
(1)Replacing Fuses
The fuses are located at the bottom on the side of the housing. Follow the
below procedure to replace fuses. Use a medium size Phillips screw driver.
1)Turn the main switch off.
2)Take the cover off the fuse box, which is located on the side of the
housing under the step. Push it up and pull it off.
3)Replace the fuse.
Take the cap off the fuse holder. (Use a medium size Phillips screw-
driver for the 8A fuse.) Take out the blown fuse and replace it with a
new one.
4)Reverse the above procedure to replace all components in their
original positions.
If the fuse blows again, this may indicate a prob-
lem with the wiring. In this case, stop using the
unit and turn the main switch off. Then contact your
local dealer or J.Morita Corp. to have the unit in
spected.
**Do not use the Maintenance Switch; this is only for use by a quali-
fied serviceman.
19. Fuses
UNIT:~ 30V PS:~ 30V EV:~ 30V AF:~ 15V CL MAIN
○ ○ ○ ○ ○ ○ ○
7A250V 5A250V 1A250V 1A250V 7A250V 8A250V
Time-Lag Fast Fast Fast Time-Lag Fast
< Fuses >
+
+
□
Maintenance-sw
**
Fuse Cover
Before replacing fuses, always turn off the main switch.
>>Otherwise, you could receive an electric shock or be affected by another type of
accident that may result in bodily injury or damage to equipment.
A blown fuse indicates a problem in the electronic circuitry, and the unit
should be inspected as soon as possible. In this case,contact your local dealer.
>>Continued use could result in accidents or malfunctions due to leaking electric-
ity, or other similar problems.
Always replace fuses with ones of the same type and capacity.
>>Using the wrong type of fuse or a substitute such as a piece of wire may result
in damage to equipment due to the wiring overheating and burning.
CHECK
Main Switch
(Green lamp goes out
when the device is
turned off.)
WARNING
CAUTION
[5]Operation 19.