12
1. Main Switch
(1)Operation
The main switch is located on the lower part of the housing.
Press it to turn the unit on. It lights up when it is turned on and all the
instruments etc. can then be operated. Press it again to turn the unit off.
(2)Save Power Function
For both safety and economy, the unit automatically enters the save
power mode if it has not been used for the set period of time.
In the save power mode,
>>The operating light goes out.**
>>The operating light and its indicator go out.
>>The power for chair operation and the gas burner is turned off.
>>The water is turned off.
When in save power mode, normal operation can be restored by pressing
any of the doctor or assistant’s operation switches, stepping of the foot
pedal, or picking up an instrument.
**: The save power mode can be turned off, or it can be set to activate
after 60 or 120 minutes. See section [5]5 3. (7) 7)User Settings” of this
manual.
* : Ootion
CAUTION
Never bump or bang the equipment or expose it to excessive vibration while
the main switch is on.
>>Doing so could cause an accident resulting in injury or damage.
Always turn the main switch off after treatment, during breaks in treatment
or whenever the unit is not in actual use.
>>Otherwise, an accident could occur resulting in injury or damage.
Turn the main switch off before using an electric scalpel in the immediate
vicinity of the equipment.
>>Otherwise, electromagnetic waves emitted by the electric scalpel could cause
the equipment to operate in a sporadic or abnormal way resulting in an accident
causing injury or damage.
[5]Operation 1.
display is off during save
power mode
Main Switch
○
Main Switch
(green lamp lights up when on)