[3]Installation
[3]Installation
The equipment must be installed by a qualified and trained professional speci-
fied by J.Morita Corp. according to the conditions specified in the Installation
Standards Manual.
>>Otherwise, there is a lisk that an accident or damage to equipment may occur.
The unit must be fastened and fixed to the floor securely as specified in the instal-
lation manual.
>>Otherwise, the chair could fall over resulting in injury or damage.
The equipment MUST NEVER be connected to an electrical supply source that does not
have the specified voltage, current, frequency and power characteristics.
Each unit must have its own devoted electrical circuit connection and should not
share a circuit with other units or other types of equipment.
>>Otherwise, short-circuiting or overheating may result in injury or damage.
Never splash the equipment with water or place it in an area where it could get
wet.
>>This will create the danger of short-circuiting or overheating, which may result in injury
or damage.
Do not install the unit in an area where chemicals are stored or where gas could
be released.
>>This would create the danger of fire or electric shock resulting in injury or damage.
Do not route the power supply cord, the earth lead or the foot control cord through
areas such as doorways and halls where they could be damaged.
>>This would create the danger of short-circuiting and electric shocks,which may result
in injury or damage.
WARNING
CAUTION
Do not tilt the equipment or place it in an unstable position; do not expose it to
excessive vibrations, bumps or shocks.
>>Doing so could cause it to malfunction or damage it.
Do not install the equipment in areas where it will be exposed to excessive atmo-
spheric pressure, humidity, temperatures, wind, direct sunlight, dust, salts or sul-
fur compounds.
>>Any of the above conditions could cause the equipment to malfunction or damage it.
The unit must be installed in a room appropriate for dental treatment.
>>If the installation location is not suitable, the unit may not only fail to perform optimally,
but it could malfunction as well.
Please take note of the following system requirements:
1)
Temperature : 10~40
℃
2)
Humidity : 30~75%RH (not be dewy)
3)
Air pressure: 700~1060hPa
>>Deviation from any of the above conditions could cause the equipment to malfunction or
result in damage.