62
When the warmer is on, all the instruments and the auto cup filler receive
a supply of warm water.
1)Warmer Switch
The switch that turns the warmer on and off is located on the basin
unit panel.
When the switch is turned on, a light above it lights up and the warmer
starts to work. The instruments will receive warm water once the heater
has had enough time to warm it up.
2)Warmer Temperature Adjustment
The warmer temperature is set at the factory before shipping, but it
can be adjusted in the following way:
a)Open the door on the basin unit.
b)Turn the temperature adjustment knob to change the temperature of
the water.
c)Close the door.
When the warmer is on, the water will not be uni-
formly warm throughout the various systems.
Watch out for places that are hot and be aware of
the temperatures of various parts of the system.
(2)Warmer
C
i
Cold
j
H
iHotj
Temperature Knob
WARNING
Whenever the warmer is turned on pay close attention to how hot the water
is. Turn the warmer off whenever it is not needed such as in the summer.
>>Otherwise, the water could get too hot and burn the patient. Also it could over-
heat and breakdown.
CHECK
[5]Operation 7.(2)
Warmer Switch
OFF
ON