48
ORTHORA 200
Instruction Manual
Possible causes
Trouble shooting
Ortholux does not move forward
Ortholux does not move back
Defective micro switch / soldered joint on
pedal 2/F2
Defective micro switch / soldered joint on pe-
dal 2/F3
Replace micro switch,
(904.0046.00)
check soldered joint
Replace micro switch,
(904.0046.00)
check soldered joint
Possible causes
Trouble shooting
9.8
Selective hose holder
Falling suction performance of cannula
Intermittant suction performance (shut-
down separation systems )
No suction at cannulas (vacuum abuts
against filter)
No suction from large suction hose
Soughing sound
Suction hose has suction in suspended
position
Full suction even though regulator valve is
in top position
Interrupted suction of any suction hose
No suction
Soiled filter in the separation system and/or
within the selective holder
Soiled replacement filter within the selective
holder and/or corner filter
Defective drain of separation system .
Clogged exterior hose
Outsized tube cross-section between selective
holder and separation system
Separation system assembled to high
Clogged vacuum supply line within filter unit
Leaking distributor valve
Loose connection between distributor hose
and tube
Soiled filter on E-valve
Soiled borehole in overflow line
Improperly attached cannula
Soiling within bellows system
Clogged valve within selective holder
Snapped off regulator hose within large suc-
tion hose
Clogged injector within selective holder
Activated motor protection switch
Defective micro switch within selective holder
Heavily soiled separation system
Clean filter
Insert new filter (320.6508.01)
Check drain and clean it if necessary
Follow the planning advises
Follow the planning advises
Pull off connection hose and clean line
Attach new suction hose (320.3032.01)
Attach distributor hose
Replace filter
Clean borehole
Clean cannula at suction hose and push it in
entirely
Unscrew lateral tappet und clean transversal
line / bellows system
Unscrew valve core and clean it (special tool
within cover of filter unit)
Shorten regulator hose and reattach it
Clean injector from top with needle. For this
purpose remove suction hose from holder for
negative pressure to result. (Needle within co-
ver of filter unit)
If necessary, disassemble selective holder and
clean injector, or replace entire suction unit.
Check power consumption
Replace micro switch, disassemble unit or
replace it entirely
Remove cover of suction engine, clean side
lines and turbine. Insert new absorption filter.