Possible causes
Trouble shooting
9. Trouble shooting
This part of the instruction is to provide assistance to technicians of
specialized dealers in particular. If you cannot localize an error by
troubleshooting, you need to commission a technician authorized by
Mikrona or Mikrona’s customer service for resolution.
9.1 Syringe
9.2 Turbine
Water drips vom turbine
Loose screw coupling
Soiled water
Inappropriately installed seal face
Hard water
Lime residues on seal face
Soft water
Pitting corrosion on seal face
Insufficient bleeding of control valve
Insufficient pressure difference air/water
Defective tappet gasket (macerated?)
Incorrect magnetic valve on control valve
Defective control valve
Defective magnetic valve
Retighten screw coupling hand-tight
Check water quality and replace water filter if
necessary
Replace water valve and tappet
(320.3069.01)
Replace water valve and tappet
(320.3069.01)
Replace water valve and tappet
(320.3069.01)
Open bleed screw
Adjust pressure values
A1 / A2
Replace tappet to water valve (320.3016.01)
Check NC or NO valve
Replace control valve 1
(320.7010.01)
Replace magnetic valve
(320.7056.01)
Possible causes
Trouble shooting
Water drips from syringe
Air / water mixing
Water leaks from couplings
No water from syringe
Loose screw coupling
Defective syringe module
Soiled valve unit within syringe module
Loose screw coupling
Defective cannula
Loose screw coupling
Defective syringe hose
Clogged cannula
Soiled valve unit within syringe module
Retighten screw coupling hand-tight
Replace syringe module
Clean valve unit or replace syringe module if
necessary
Retighten screw coupling hand-tight
Replace cannula
Retighten screw coupling hand-tight
Replace syringe hose
Replace cannula
Clean valve unit or replace syringe module if
necessary