13
Instruction Manual
5. Installation at customer
5.1
Chair installation
Practice installation must be carried out in accordance with the re-
quirements stated in the installation manual.
Plan prints in scale 1:1 (320.8002.01)
6. Connections
6.3
Air connection
Air connection ground:
R 3/8 (external thread)
Air connection device:
G 1/4 (external thread)
Above ground:
10 mm
Air input:
min. 5 bar
Air passage:
max. 75 l/min
The scope of delivery includes the low-pressure line 906.0137.01 with
R 3/8 (internal thread) and G 1/4 (internal thread).
6.4
Water connection
Water connection ground:
R 3/8 (external thread)
Water connection device:
G 1/4 (external thread)
Above ground:
10 mm
Water input:
Domestic water pressure,
min. 2.5 bar
Water passage:
max. 1.5. l/min
Water quality:
Drinking-water
Drain connection:
Ø 36 mm (inner diameter)
(HDPE pipe according to
DIN 8074 T2)
Above ground:
10 mm
Outflow rate:
max. 6.5 l/min with free outflow
Outflow rate:
max. 3.5 l/min with wet
aspiration
Inclination drain pipe:
1 % (10 mm per meter)
Water filtration on device:
80 µm
Ph-value: 7.2-7.8
Water hardness:
7-11°dH / 13-19°fH
If the water is harder than indicated, it is necessary to install a water
softening device. If the water is softer than indicated, this may lead to
growth of algae.
The scope of delivery includes the low-pressure line 906.0137.01 with
R 3/8 (internal thread) and G 1/4 (internal thread).
6.1
Electrical connection
Electrical supply line:
3x1.5 mm
2
Free end above ground:
500 mm
Input voltage:
100 V AC / 120 V AC / 230 V AC
country-specific
Apparent power:
1900 VA
Power input unit:
300 VA (power socket 1'400 VA)
Frequency:
50 / 60 Hz
Fuse:
6.3 A
6.2
Aspiration connection
Aspiration control line:
3x1.5 mm
2
Aspiration connection:
Ø 36 mm (inner diameter)
(HDPE pipe according to
DIN 8074 T2)
Above ground:
10 mm
Aspiration air rate:
300-350 l/min
(main aspiration hose)
Depression:
ca. 180 mbar (hPa)
6.5
Transport and storage conditions
Temperature:
-20°C up to +60°C (-4°F up to 140°F)
Relative humidity:
10% up to 75%
Air pressure:
500hPa up to 1060hPa
6.6
Operating conditions
Ambient temperature:
15°C up to +35°C (59°F up to 95°F)
Relative humidity:
30% up to 75% (no condensation)
Air pressure:
860hPa up to 1060hPa