35
Instruction Manual
7.17.3 ORTHOLUX 200
General information
For information on technical features as well as on operation and ap-
plication of this examination lamp please consider the specific instruc-
tion and maintenance manual which comes with the lamp.
Description
The ceiling-mounted ORTHOLUX 200 lamp head is activated by a
3-function pedal in the treatment unit (see 7.2.2 Special functions).
Besides of switching the lamp on and off, the lamp head can also be
swiveled by 30° (adjustment of light field position). The lamp is fed and
controlled from the treatment unit.
Commissioning and use
Only use instruments in perfect operating condition. Immediately stop
any work if intermittences or other signs of product failure occur.
Cleaning
Do not clean the protective cover while the lamp is activated and let
the halogen lamp and the reflector cool down sufficiently before remo-
ving the protective cover.
Unscrew the central attachment bolts of the protective cover (counter-
clockwise) and remove the protective cover.
The protective cover is made of acrylic glass and easy to clean. Clean
stained surfaces by using warm water with a non-abrasive household
cleaner and a soft tissue or cloth (cotton). Avoid under any circum-
stances to use a dry or dirty cloth as this may lead to scratches on
the acrylic glass. Never use solvent-containing products or cleaning
agents (e.g. benzene, alcohol etc.). These agents can damage and
cloud the surface.
Possible scratches can be removed using Acrylic glass polish.
Disinfection
Use a moist cloth to disinfect the case surface.
Do not spray any disinfectant solution on the device.
Replacing the halogen lamp
Please follow the corresponding operating and maintenance instruc-
tions which come with the lamp. Do not replace the halogen lamp
while it is activated (Main switch at treatment unit in OFF position, see
7.1 Main switch) and let the halogen lamp and the reflector cool down
sufficiently before replacing it.
Adjusting the focus
Please follow the corresponding operating and maintenance instruc-
tions which come with the lamp.
Maintenance
No specific maintenance procedures are intended for this devise.
Under any circumstances avoid using any lubricants as this my cause
irreparable damage to the ORTHOLUX 200 ceiling lamp.