24
ORTHORA 200
Instruction Manual
7.9
Turbine connection
General information
For information on technical features as well as on operation and
application of the turbine please consider the specific instruction ma-
nuals which come enclosed in the instrument and motor packaging.
Please service and handle the turbine hand pieces as specified in the
maintenance instructions of the specific manufacturer.
Description
Turbine device with 4VML LUX standard connection 1.5 to 3.0 bar for
dental application with integrated spray and light.
Commissioning
Never attach an instrument while the turbine hand piece is running.
Only use instruments in perfect operating condition. Immediately stop
any work if intermittences, extensive vibrations, abnormal heating or
other signs of instrument failure occur.
For use, remove the instrument from the rack (micro switch will be
activated with a rack circuit installed). The display shows the specific
instrument as well as the speed level. Start the instrument by pressing
the function pedal. Change speed and switch on the spray by using
the function pedal (7.2.3. or 7.2.4).
Sterilizing the turbine hand piece
The turbine hand piece (which is not included in the scope of delivery
of Mikrona Technologie AG) should be serviced and handled accor-
ding to the manufacturer’s maintenance instructions. For information
on technical features please consider the specific instruction manuals
which come enclosed in the instrument and motor packaging.
Disinfection
For disinfection use a clean damp cloth and disinfect the instrument
tube. Only use disinfectants recommended by the manufacturer.
Never submerge the device in a disinfectant solution.
Do not insert the device into an ultrasonic bath.
Default settings
Drive air
max. 2.5 bar
Spray air
2 bar
Water
ca. 50 ml/min
4-function pedal control 09 (ST09) units with serial numbers > OT1.02000
X
_
X
_
Level 1
1,5 to 2,0 bar
10 to 25%
Level 2
2,0 to 2,5 bar
25 to 40%
Level 3
2,5 to 3,0 bar
40 to 99%
Level 4
1,5 to 3,0 bar
10 to 99%
5-function pedal control 09 (ST09) units with serial numbers > OT1.02000
X
_
Level 1
1,5 bar
10%
Level 2
2,0 bar
25%
Level 3
2,5 bar
40%
Level 4
3,0 bar
99%
5-function pedal control 01 (ST01) units with serial numbers < OT1.01999
Bien Air Kavo
Level 1
1,5 bar
10,8 V
15 V
Level 2
2,0 bar
13 V
18 V
Level 3
2,5 bar
15 V
20 V
Level 4
3,0 bar
19 V
25 V