31
Instruction Manual
7.16 Polymerization lamp
MINI LED ORTHO
General information
For information on technical features as well as on operation and
application of this polymerization lamp please consider the specific in-
struction manual which comes with the instrument. Please handle the
hand piece as specified in the maintenance instructions of the specific
manufacturer.
The light rays radiating from this device can be dangerous and should
never be directed towards the eyes, not even when wearing protective
goggles. Only direct the light towards that part of the oral cavity which
requires treatment. Please follow the safety measures recommended
by the manufacturer.
Description
This polymerization lamp has been designed exclusively for dental ap-
plication. It radiates a visible blue light with a wavelength of 420 to 480
nm for the photopolymerization of dental material.
Commissioning and use
Only use instruments in perfect operating condition. Immediately stop
any work if intermittences, abnormal heating or other signs of instru-
ment failure occur.
To start the polymerization cycle briefly press the ON/OFF button. A
first beep sounds.
The end of the cycle is indicated by a second beep.
The cycle can be interrupted at any time by slightly pressing the ON/
OFF button.
Cleaning and sterilizing the light conductor
Remove the light conductor and clean and sterilize it before using it on
the next patient.
Sterilization occurs in an autoclave at 134°C.
Disinfection
For disinfection of hand piece and hose use a clean damp cloth and
disinfect the case surface. Only use disinfectants recommended by
the manufacturer.
Never submerge the device in a disinfectant solution.
Do not insert the device into an ultrasonic bath.
Maintenance
No specific maintenance procedures are intended for this devise.
Under any circumstances avoid using any lubricants as this my cause
irreparable damage to polymerization lamp.