37
Instruction Manual
7.20 Autonomous water supply
General Information
Optionally, instruments can be supplied with water from an autono-
mous water container. Fresh water must be changed on a daily basis.
Fresh water can also be treated with disinfection additives
(e.g. Alpron) for process water.
Description
The water container is supplied with compressed air from the treat-
ment unit.
The container (water bottle) comes in the size of 0.8 L and 1.2 L.
Commissioning and use
1 = Bottle system ON (the bottle is pressurized)
2 = Bottle system OFF (the bottle is depressurized)
A Depressurize the water bottle (toggle switch on 2 / OFF).
B Unscrew the water bottle (clockwise).
C Fill the water bottle with tap/drinking water (if need be for an opti-
mal hygiene of the process water e.g. Alpron can be added).
D Attach the water bottle with about 1 turn to the mounting head
(counter-clockwise).
E Pressurize the water bottle (toggle valve on 1 / ON).
Tighten the water bottle until no more leakage noise can be heard.
F Start the instrument.
G Always depressurize the system when refilling tap/drinking water
(toggle switch on 2 / OFF).
Only use products in perfect operating condition. Immediately stop
any work if intermittences or other signs of product failure occur.
Disinfection
In idle periods or over night, rinse and sterilize the autonomous water
supply and the water-bearing instrument tubes with a disinfectant solu-
tion (e.g. Bilpron)
A Depressurize the water bottle (toggle switch on 2 / OFF).
B Unscrew the water bottle (clockwise).
C Empty the water bottle from tap/drinking water.
D Fill the water bottle with a disinfectant solution.
(Basic liquid approx. 50ml, additionally approx. 50ml per instrument.)
E Attach the water bottle with about 1 turn to the mounting head
(counter-clockwise).
F Pressurize the water bottle (toggle valve on 1 / ON).
Tighten the water bottle until no more leakage noise can be heard.
G Start the instrument and rinse the tubes.
H Let the disinfectant solution remain in the water-bearing tubes du-
ring the idle period.
I
Before the next use on a patient the disinfectant has to be drained and
all the water-bearing tubes to the instruments have to be thoroughly
rinsed with tap resp. drinking water (approx. 50ml per instrument).
Maintenance
No specific maintenance procedures are intended for the autonomous
water supply. Any repairs such as making adjustments or changing
settings are only to be carried out by qualified technicians or Mikrona’s
customer service.
1
2
Remove scratches with a soft steel wool or Scotch (Scotch-Brite Hand
Pad Nr. 7447), rubbing in direction of the surface’s grain.
Disinfection
Use a moist cloth and disinfect the surface with a standard disinfectant
(e.g. Dürr Surface Disinfection FD 322 / FD 350).
Maintenance
No specific maintenance procedures are intended for the suspended
table. Any repairs such as making adjustments or changing settings
are only to be carried out by qualified technicians or Mikrona’s custo-
mer service.