74
Operating Instructions - CPG2500
Precision Pressure Indicator
CPG2500
10.6 Calibration Data
The Calibration Data Application is where the calibration data for each transducer is stored and
amended. The Serial number (S/N), Zero offset (Zero) and Span offset (Span) can be seen in this screen.
The date of calibration, the calibration interval and the certificate number can be entered by pressing
the corresponding button, then saved by pressing the Check [
] button. To revert back to the factory
calibration, press the “Restore Factory Cal” button. To view the calibration data for each transducer press
the “Channel” button at the top and choose a transducer from the resulting channel selection menu.
Figure 10.6 - Calibration Data
10.7 One Point Cal Application
A single point calibration (usually a zero point calibration) of each transducer installed can be
accomplished in “One Point Cal” Application. The transducer channel is chosen by pressing the
“Channel” button at the top of this screen, and then selecting from the list of installed transducer
channels. For gauge pressure simply expose the reference and the pressure port of the transducer to
atmospheric pressure and then press the “New Value” button and enter zero (0) using the keypad. For
an absolute transducer apply a known reference pressure between 600 mTorr absolute and 20% of the
selected transducer’s span to the pressure port of the transducer, press the “New Value” button and then
enter the reference pressure (known true pressure) using the keypad. If you want to save the value in the
sensor, press Save.
Notice the reference symbol [ ],
this is a reference indication giving
a constant reference point for the
level of the internal sensor.
Figure 10.7 - One Point Cal Application
Содержание CPG2500
Страница 14: ...14 Operating Instructions CPG2500 Precision Pressure Indicator CPG2500 NOTES ...
Страница 47: ...Operating Instructions CPG2500 47 Precision Pressure Indicator CPG2500 NOTES ...
Страница 61: ...Operating Instructions CPG2500 61 Precision Pressure Indicator CPG2500 NOTES ...
Страница 67: ...Operating Instructions CPG2500 67 Precision Pressure Indicator CPG2500 NOTES ...
Страница 69: ...Operating Instructions CPG2500 69 Precision Pressure Indicator CPG2500 NOTES ...
Страница 85: ...Operating Instructions CPG2500 85 Precision Pressure Indicator CPG2500 NOTES ...