48
Operating Instructions - CPG2500
Precision Pressure Indicator
CPG2500
7 Remote Operation
Use the screens in Section 6.4.5 Remote Application to set the operating parameters for the instrument
command set, Ethernet, Serial (RS-232) and IEEE-488 (GPIB) information.
7.1 Command Set
Command Set button – Users’ can select which model remote protocol they would like to emulate for
simulation and testing purposes. Selections may include the following or may be added per customers’
specifications:
• Mensor (default)
• SCPI WIKA (The SCPI WIKA mode emulates the WIKA command set in SCPI format.)
7.2 IEEE-488
IEEE-488 address button – Allows the user to set the GPIB address by inputting a numeric value utilizing
the touch screen.
7.2.1 IEEE-488.2 Commands
Command or Query
Response / Function
*IDN?
Returns identification string
*RST
Reset to known state (psi)
*TST?
Returns 1
*OPC
Operation completed
*WAI
Returns operation completed state
*CLS
Clear status and error queue
*ESE
Enable status event
*ESE?
Returns enable status even value
*ESR
Event status register
*ESR?
Returns even status register value
*SRE
Service request enable
*SRE?
Returns service request enable value
*STB?
Returns status byte
Содержание CPG2500
Страница 14: ...14 Operating Instructions CPG2500 Precision Pressure Indicator CPG2500 NOTES ...
Страница 47: ...Operating Instructions CPG2500 47 Precision Pressure Indicator CPG2500 NOTES ...
Страница 61: ...Operating Instructions CPG2500 61 Precision Pressure Indicator CPG2500 NOTES ...
Страница 67: ...Operating Instructions CPG2500 67 Precision Pressure Indicator CPG2500 NOTES ...
Страница 69: ...Operating Instructions CPG2500 69 Precision Pressure Indicator CPG2500 NOTES ...
Страница 85: ...Operating Instructions CPG2500 85 Precision Pressure Indicator CPG2500 NOTES ...