20
Operating Instructions - CPG2500
Precision Pressure Indicator
CPG2500
5.3 Mounting
The instrument can be set up on a desk top or it can be rack-mounted. Rack mount hardware is optional
on the CPG2500 (see Section 5.2 Dimensions and Section 8, Options).
The special sensors used in the CPG2500 are relatively insensitive to tilt and vibration. However to fur-
ther assure stability and accuracy, avoid mounting the instrument on surfaces subject to excessive motor
or machinery vibration.
5.4 Rear Panel
Two slots are available on the rear of the CPG2500. These slots can be filled with a removable transducer
or a remote transducer interface connection. Gauge Transducers will have a reference and a pressure
port. Absolute transducers will have a plug in the reference port. In the upper right corner is a hose barb
fitting which is connected to the baro ref if installed. Transducers with analog output will have an electrical
connection. Positioned on the left side is a remote transducer connection, the RS-232, Ethernet, IEEE-
488, USB device connections for communication, the USB host connection and the 12 VDC power input.
Figure 5.4 A - Two removable gauge transducers installed
Figure 5.4 B - Two remote transducer interface connections installed
Содержание CPG2500
Страница 14: ...14 Operating Instructions CPG2500 Precision Pressure Indicator CPG2500 NOTES ...
Страница 47: ...Operating Instructions CPG2500 47 Precision Pressure Indicator CPG2500 NOTES ...
Страница 61: ...Operating Instructions CPG2500 61 Precision Pressure Indicator CPG2500 NOTES ...
Страница 67: ...Operating Instructions CPG2500 67 Precision Pressure Indicator CPG2500 NOTES ...
Страница 69: ...Operating Instructions CPG2500 69 Precision Pressure Indicator CPG2500 NOTES ...
Страница 85: ...Operating Instructions CPG2500 85 Precision Pressure Indicator CPG2500 NOTES ...