64
Operating Instructions - CPG2500
Precision Pressure Indicator
CPG2500
A plate with a D-Sub connector can be supplied with the instrument or after the initial purchase. This
plate (See picture in section 5.4 Figure 5.4 B) can be installed in the place of a removable transducer to
provide an additional connection for an external transducer (See section 8.4). When replacing a remov-
able transducer with the plate, the internal ribbon cable that was connected to the removable transducer
should be connected to the inside connector on the plate.
8.6 Analog Output
The analog output option is available on all internal Transducers and on the CPT6100 external trans-
ducer. It can be configured at the time of manufacture to have voltage output that is directly proportional
to the pressure input.
8.6.1 Specifications
Resolution: 0-1 VDC
1 part per 80,000 minimum
0-5 VDC
1 part per 400,000 minimum
0-10 VDC
1 part per 800,000 minimum
Update Rate: 150mS
Uncertainties: 0-1 VDC
0.010% FS
0-5
VDC 0.010%
FS
0-10
VDC
0.010%
FS
Minimum input impedance of the device measuring the analog output is 1 Megaohm.
8.7 Rack Mount Kit
The CPG2500 can be supplied with a single or dual 19” rack mount kit.
Dual Rack Mount Single Rack Mount
8.8 Fittings
A variety of fittings are available at the time of order on sensors with a max range of ≤ 6015 psi (410
bar), or the CPG2500 can be ordered without adaptor fittings. Available fittings are shown in Section 9.2
- Spare Parts. An Autoclave F250C/ HIP HF4 is integral in all transducers with full scale ranges above
6015 psi.
8.9 Remote Calibration
A remote calibration sled, power supply, cable and software are available to calibrate the optional baro-
metric reference. A cable / power supply and software are available to remotely calibrate the internal
transducers outside of the CPG2500. In both cases a PC with a serial connection is required. Both kits
are supplied with an instruction manual.
Содержание CPG2500
Страница 14: ...14 Operating Instructions CPG2500 Precision Pressure Indicator CPG2500 NOTES ...
Страница 47: ...Operating Instructions CPG2500 47 Precision Pressure Indicator CPG2500 NOTES ...
Страница 61: ...Operating Instructions CPG2500 61 Precision Pressure Indicator CPG2500 NOTES ...
Страница 67: ...Operating Instructions CPG2500 67 Precision Pressure Indicator CPG2500 NOTES ...
Страница 69: ...Operating Instructions CPG2500 69 Precision Pressure Indicator CPG2500 NOTES ...
Страница 85: ...Operating Instructions CPG2500 85 Precision Pressure Indicator CPG2500 NOTES ...