12.3 Warning Displays
(continued)
A counting error has been detected in
Sampling Device mode. Verify that
tubes are in correct position and
order. (See Section 5.9 for more
detail.)
OT wash device was touched while
sampling device was running. See
Section 5.9 if emergency sample
analysis is needed.
Incorrect authorization code was
entered. See calibration section in
User’s Manual for entry of correct
authorization code for calibration, or
contact local distributor or
authorized service technician for
service related authorization codes.
Authorization Code and Installation Warning Displays
No authorization code was entered.
See calibration section in User’s
Manual for entry of correct
authorization code for calibration, or
contact local distributor or authorized
service technician for service related
authorization codes.
Reagent level sensors must be
removed from the reagent containers
when emptying the system. Verify
that both level sensors have been
removed.
Instrument has detected liquid in the
system. The empty cycle must be
run prior to a fill cycle. Run the
Empty function to remove any extra
liquid remaining in the system then
fill the instrument with reagents.
Continued on next page