83
11.2 Short List of Specifications
Specifications (Short)
Measuring principle, RBC, WBC, PLT
Impedance
Measuring principle HGB
Photometer, Cyanide free method 535nm ±5nm
Programmable WBC Discriminator
Yes
Sampling system
Closed shear valve
Parameters reported
RBC, MCV, HCT, PLT, MPV, HGB, MCH, MCHC,
WBC, RDW%, LYM abs, MID abs, GRAN abs, LYM%,
MID%, GRAN%
Size distributions printed for
RBC, PLT and WBC diff.
Aspirated blood volume (Open Tube)
< 110 µl
Aspirated blood volume (Cap Piercer)
< 250 µl
Aspirated blood volume (Autoloader)
< 300 µl
Blood volume Micro Pipette Adapter (MPA)
20 µl
Pre-diluted mode
1:200 to 1:300 using min. 20 µl
e.g. 20 µl to 4.5 ml diluent (1:225)
TFT-LCD Display
Graphical color touch screen, 240 columns x 320 rows
Keyboard
Virtual incorporated keyboard (External keyboard option)
Number of Samples per hour (Open Tube)
> 60 samples
Number of Samples per hour (Cap Piercer)
> 45 samples
Number of Samples per hour (Autoloader)
> 43 samples
Sample display time (Open Tube)
< 50 seconds
Printer
External, Seiko DPU 411/12 and 414, IBM proprinter /
Epson compatible, HP PCL 3 and 5 protocol compatible,
USB printer
Control sample memory capacity
> 1000 control samples
Sample memory capacity
> 1000 samples
QC capabilities
Mean, SD, CV, Levey-Jennings plots and X-B with
>10,000 samples history
HGB correction on high WBC counts
Yes
Warning flags on parameter abnormalities
Yes
Floating discriminator RBC/PLT
Yes (position printed)
Mathematical 3-part diff. WBC calculation
Yes
Automatic HGB blank on each sample
Yes
Carry over
< 1 %
Barcode reader input
Yes
Serial output
Yes (Conformed to standard EN 60950)
Main Voltage
100 – 240 V AC
External Power Adapter 24 V DC
Mains voltage tolerances
±15 %
Power consumption
Max 100VA
Power consumption (stand-by)
Max 20VA
Frequency
50 / 60 HZ
Built-in test / adjustment programs
Yes
Temperature
64 - 90°F (18 - 32°C)
Humidity (noncondensing)
Up to 80%
Dimensions (Basic/Standard/Closed Tube)
HxWxD = 16 x 11.5 x 18 inches (410 x 290 x 460 mm)
Dimensions (Autoloader)
HxWxD = 17 x 13 x 18 inches (430 x 330 x 460 mm)
Instrument weight (Basic/Standard/Closed Tube) ≤ 40 lbs (18 kg)
Instrument weight (Autoloader)
≤ 48.5 lbs (22 kg)
Diluent Consumption
Approximately 22 ml per analysis cycle.
Lyse Consumption
Approximately 4.5 ml per analysis cycle.
Содержание M16C-US