99
12.5 Troubleshooting Other Issues
Description
See Troubleshooting Flowchart in Appendix A for other possible issues that
may arise. Areas on Flowcharts highlighted in dark grey should only be
performed by service technician or authorized personnel.
Indication Error
Codes
Indications error codes are specific instrument situations that in most cases
need the attention of the operator or might need service action.
The three number indications usually occur after the two number
indications. For example, an indication 302 will be displayed due to
interference with an OT analysis. It states that the OT cycle was
aborted.
The first indication display is the most important as it describes the
issue and how to solve the problem. The three digit indication after a
two digit one is added information for the user.
In most cases, the instrument is stopped and the operator has to confirm
with [OK] to continue. Once [OK] is pressed and instrument returns to
display menus, user should repeat previous actions again (e.g. re-
analyze sample, printing results, etc.)
If indication error appears again or a three digit indication was
displayed as the first indication message, contact local distributor or
authorized service technician.
Indication Series Description
1 - 19
Indication series for auxiliary errors like battery faults or similar.
20 - 29
Indication series for ‘Liquid’ errors.
30 - 39
Indication series for Communication errors between the PCBs (CAN
bus).
40 - 49
Indication series for Printer and serial output errors.
50 - 59
Indication series for General Memory errors.
60 - 69
Indication series for EEPROM/HPC (High Performance Controller)
errors.
70 - 79
Indication series for Shear Valve problems.
80 - 89
Indication series for Cap Piercer errors (Closed Tube Adaptor).
90 - 99
Indication series for Sampling device errors.
100-255
Indication series for internal hardware and software problems, and
messages during subboard firmware upgrades.
300 -399
Indication series for cycle aborted indication numbers.
Содержание M16C-US