50
5.9 Analyzing the Sample (Autoloader)
(continued)
Emergency
Sample Analysis
Emergency (STAT) samples can be analyzed after the Sampling Device has
been started or during Sampling Device ID entry. There are several ways to
analyze an emergency sample.
Step
Action
1
Emergency sample can be analyzed through OT, pre-dilute, or MPA mode.
Press [PAUSE], wait for [NEW SAMPLE] button to highlight green, and
then analyze sample in preferred mode.
There may be a slight delay after pressing [PAUSE] button before
emergency sample can be analyzed. This is because analyzer will
complete the counting cycle of the last sample run on sample wheel
before continuing with emergency sample analysis.
When emergency sample is complete, press [CONTINUE] to restart
sampling in next position on the wheel.
2
Emergency sample can also be analyzed using the sample wheel.
Press [STOP], unlock sample wheel and place emergency sample in
Position 1 or 21.
If a sample is already occupying Position 1 or 21 and has already been
analyzed, remove sample and place emergency sample in its place.
If emergency sample has a barcode for ID number, align barcode
correctly, lock sample wheel and press [CONTINUE].
See Editing Sample ID number is manual entry of sample is desired, and
lock sample wheel and press [CONTINUE].
Analyzer will automatically analyze emergency sample and then
continue sampling where it left off prior to pressing [STOP] button.
Note
DO NOT press [START] after sampling device has been paused or stopped
unless operator wants to rerun all samples on wheel.
Control Sample
Analysis
If analyzing samples using the Autoloader mode it is recommended to also
run daily control samples using the sample wheel.
Step
Action
1
Follow instruction in Section 6 for control handling and assay sheet input.
2
Firmly press capped end of control sample into control tube adapter.
3
Load the control sample by placing the adapter towards the outer edge of the
sample wheel and fitting it into Position 1 for all tubes except Sarstedt. Place
Sarstedt control sample in Position 40.
Position control tube barcode facing TOWARDS analyzer and centered
in slot.
If using all three levels of control, add adapters to all levels of controls
and fit them into Positions 1, 2, and 3.
4
Following instruction below for Starting Sampling Device.
Continued on next page