![Medonic M16C-US Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/medonic/m16c-us/m16c-us_user-manual_1762714094.webp)
93
12.3 Warning Displays
Warning
Displays
Warning displays appear after a function has been performed incorrectly or to
inform the operator that further action is needed to complete the desired task.
The warning display describes the situation and instructs the operator on next
step or function to resolve issue.
System Power Down Warning Displays
The system has been switched off
for a long time period. The
instrument has been powered down
with all valves open and filled with
liquid. Empty and refill the system
with reagents, and perform a
background count.
The system was switched off
incorrectly. Perform a prime to
prepare the system for analysis. Check
method for correct instrument power
down procedure.
The system was manually switched
off with system emptied of reagents.
Fill the instrument with reagents to
prepare for analysis or exit if only a
search of instrument menus is
needed.
The instrument has been switched
off with power down function
before power was switched off.
Perform a power up to prepare the
reagent system for analysis.
The system was powered down with
liquid in system and has been unused
for long period of time. Perform the
cleaning procedure according to
cleaning kit instruction. Perform a
background check.
The regular 12 hour wash has failed.
Make sure that reagent containers
are filled and the detectors are
inserted correctly.
Continued on next page
Содержание M16C-US