8.3 Six (6) Month Cleaning
Description
To increase the life of internal tubing in the instrument, the following cleaning
procedure is strongly recommended.
Cleaning
Procedure
Press [ADVANCED] from Main menu, then press [MAINTENANCE], and
then press [CLEANING MENU] to enter the Cleaning Menu.
Follow the instruction for the Boule Cleaning kit to clean the instrument.
(Instructions for use are supplied with the Boule Cleaning kit solutions).
The Six Month Cleaning procedure takes approximately one hour and 15
minutes to complete.
Figure 8.1 Figure 8.2 Figure 8.3
Boule Cleaning
Kit
The Boule Cleaning Kit contains the following items:
Enzymatic cleaner (Bottle #1)
Hypochlorite (2%) (Bottle #2)
Detergent cleaner (Bottle #3)
Cleaning
Interval
Depending on daily sample analyses, it is recommended that the following
cleaning intervals be followed:
Less than 50 samples/day = every six months
More than 50 samples/day = every three months
100 – 200 samples/day = every month