38
5.2 Startup Sequence
(continued)
Step
Action
5
When complete the background count results are displayed. If the results
are acceptable (see table for accepted background values according to
section 5.3)
, scan in the barcode on control vial and follow directions on
the display to begin the second step of the startup sequence.
Note
If the background count results have a H (high) indicator press [RERUN]
and follow the screen instructions to analyze background count again.
Figure 5.3 Figure 5.4
6
When complete the control results are displayed. If control results are
acceptable, press [RERUN] to run next level of control. The startup sequence
is complete when all control results are acceptable. Press [ANALYZE
SAMPLES] to go to the main screen, and follow instructions in the
following sections to analyze samples.
Note
If control sample results have a H (high) or L (low) indicator press
[RERUN] to analyze control sample again.
Figure 5.5 Figure 5.6