2.4 Reagent Installation
(continued)
Step
Insert
3
The reagent level sensors into the corresponding reagent containers.
Figure 2.10
Waste
Connect the waste tubing to the analyzer. Place the other end of the waste
tubing directly into the drainage system or into a waste container, following
local regulations. See Section 8.8 for Disposal information.
Caution
The end of the waste tubing must be at a lower level than the instrument itself.
Not following this may lead to improper instrument functions and/or waste
liquid flowing backwards into the instrument.
Mandatory Action
Always use protective gloves when working with the waste container and the
waste tubing.
Fill System
For initial fill of analyzer, plug in analyzer and turn On/Off switch to ON.
Press [EXIT] button upon display of Fill prompt, and follow the instructions
below to fill analyzer.
Continued on next page