13
2.2 Installation Checklist and Menu
(continued)
Step
Action
2
Press Step 2 [ENTER REAGENT BARCODES].
Scan barcode 1 and then barcode 2 on the Diluent container. (
Press and hold
the ACTIVE or ON button each time a barcode is scanned.)
o
If using a Combination pack, following instruction for scanning in
Diluent container.
If using single containers of Diluent and Lyse press [ENTER ANOTHER
BARCODE] and scan barcode 1 and then barcode 2 on the Lyse container.
Press
[EXIT] to return to Reagent Barcode Input screen and then press
[EXIT] again to return to the Installation Menu.
Note
After reagents are scanned, then loosen reagent container caps, remove
factory seals, and place reagent level sensors in respective containers.
Figure 2.4 Figure 2.5
3
Press Step 3 [ENTER CONTROL BARCODES] to enter assay value ranges into the
system for the lot of Control being used.
Scan barcodes 1-9, in that order, for each control level.
Once accepted press
[EXIT] to return to Installation Menu.
4
Press Step 4 [PERFORM FILL SYSTEM] to fill system with reagents. This cycle
will last for approximately 3 minutes.
Figure 2.6 Figure 2.7
Optional
Press Step 5 [GO TO STARTUP]. See Section 5.2 for details on guided startup
sequence.
Содержание M16C-US
Страница 1: ...Medonic M series...