62
7.2 Calibration
(continued)
14
Authorized operator can update or change calibration factor by inputting
the Authorization Code [
2576
].
Figure 7.3
15
Perform steps 9-12 for RBC, MCV, PLT, HGB, WBC, and MPV
parameters. To move to the next parameter press [NEXT].
16
It is recommended to not change preset calibration factors for RDW%. If
necessary, please contact local distributor or Boule service technician for
procedure.
17
Once parameters are calibrated, press [EXIT] and a screen will be
displayed asking operator if a calibration report is wanted, [SEND],
[PRINT], or [EXIT] can be selected. It is recommended that calibration
reports be printed and archived in case it may be needed for future
reference.
Figure 7.4
18
It is recommended to run controls after calibration to verify that all
parameters have been calibrated correctly. See section 6.1 to perform QC.
Capillary Device
Calibration
To calibrate MPA follow Steps 1-17 above except select
[CALIBRATION] and then choose [CAPILLARY DEVICE] instead of
Whole Blood calibration in Step 6 and use MPA mode for analysis. (See
Section 5.7 for details on capillary device sample analysis.)