54
Section 6: Quality Control (QC) and Blood
Control Memory
Section Overview
Introduction
The Medonic M-Series is equipped with a QC memory capable of
displaying and printing Xb and Levey Jennings plots.
Contents
This section contains the following topics:
Topic
See Page
Quality Control (QC)
54
Levey-Jennings Plots
57
Initialization and Use of Xb Function
58
6.1 Quality Control (QC)
Introduction
This section describes the procedures to be performed for running control
samples.
QC Menu and
Assay Value Input
Follow the instruction below to access the QC menu and to input
Control/Calibrator Assay Values from the Assay sheet.
Step
Action
1
Enter the QC menu by pressing [QC] from the menu tab.
2
Press [ENTER CON/CAL].
3
Refer to the Assay sheet for instructions on how to input Assay Values.
(These pages are delivered with authorized Boule controls).
Figure 6.1 Figure 6.2
Note
12 different Assay Lots from Boule can be stored simultaneously.
When entering a new Assay Lot, the previously scanned Assay Lot will
be removed in a chronological order starting with the first entered Lot.
Continued on next page