95
12.3 Warning Displays
(continued)
Barcode Warning Displays
No more space is available to scan
in new assay values. Follow the
recommendation or manually delete
all the controls with same ID, to
free space for scanning the new
Control lot. (See Section 6.1 for
more detail.)
Assay Value Input failed. The Assay
sheet or order of scanning in the
barcodes may have been incorrect.
Verify that setups on the instrument
match the required setup for the
barcode reader. (See Section 4.3 and
6.1 for more detail.)
Reagent barcode scanning failed.
Barcode printing or order of
scanning in the barcodes may have
been incorrect. Verify that setups on
the instrument match the required
setup for the barcode reader. (See
Section 4.3 and 4.4 for more detail.)
Open Tube Warning Displays
The instrument was unable to wash
the Open Tube probe (needle).
Verify that tube is removed and
wash device is in correct position,
then perform OT Wash.
The instrument was unable to wash
the Open Tube probe (needle). Verify
that tube is removed and wash device
is in correct position. It is
recommended that background count
is performed before next sample
analysis.
The instrument is unable to wash the
Open Tube probe (needle). Verify
that tube is removed and wash
device is in correct position. It is
recommended that background count
is performed before next sample
analysis.
Continued on next page