Ca++ is replaced with Cl- when Cl- mode is selected.
CONFIGURATION
Select
CONFIGURATION
to set analyzer options. Select from:
PO2/PCO2/BP
,
Na+/K+/Ca++
,
Hct
,
Hb
,
CAL
FREQUENCY
,
AUTO STANDBY
,
and
TEMPERATURE
.
PO2/PCO2/BP
Change the measurement units (select mmHg or SI units) for PO2,
PCO2 and barometric pressure. Results for SI units are shown in KPa.
Na+/K+/Ca++ or Cl-
Change the measurement units (select mmol/L or mmEq/L units).
HCT
Change the measurement units (select % or V/V) for hematocrit.
Hb
Change the measurement units for Hb. Select mmol/L, g/L or g/dL.
BC Checksum
Turn the feature ON or OFF. When the feature is ON, the analyzer
has the uses the last digit in the barcode to perform a barcode check-
sum calculation.
CAL FREQUENCY
Press the number keys to select and change an option. Press YES to
CONFIRM?
all configuration settings.
Select the time interval between automatic two-point calibrations.
Select from: 2 hours, 4 hours, 8 hours
1 PO
2
/ PCO
2
/ BP
mmHg
2 Na
+
/
K
+
/
Ca
++
mmol/L
3 Hct
%
4 Hb
g/dL
5 BC CHECKSUM OFF
CONFIRM ?
CONFIGURATION 1/2
1 CAL FREQUENCY
8 HRS
2 AUTO STANDBY
OFF
3 TEMPERATURE
C
4 Rili-BAK OFF
5 AMR
OFF
CONFIRM ?
CONFIGURATION 2/2