t r o u b l e s h o o t i n g
MILLIVOLT RANGE GUIDELINES:
Calibrant A
Calibrant B
Calibrant C
pH
(mV)
-350 - 350
-320 - 380
PCO2
(mV)
-150 - 150
-135 - 165
PO2
(mV)
150 - 800
less than 60
Hct
100 - 250
300 - 600
Na+
(mV)
100 - 200
85 - 185
K+
(mV)
100 - 200
115 - 215
Ca++
(mV)
30 - 130
40-140
40 - 140
Cl-
(mV)
30 - 80
90 - 190
•
Replace any electrode with mV’s out of range.
The Na+, K+, Ca++, Cl-, and pH electrodes are dependent on the reference elec-
trode mV readings. If the electrode mV readings are out of range, first verify that
the reference electrode is within its use period (refer to the Replacement Schedule
in Chapter 11).
•
In the case of multiple electrode mV failures, contact your EasyStat
dealer.
s
Continued electrode error messages may indicate elec-
trode failure.
•
Check the electrode millivolt readings by selecting
SENSOR
STATUS
, then
ELECTRODE mV
’
s
from the
DIAGNOSTICS
menu.
•
Check both Calibrant A, B and C mV’s. Wait 20 seconds for the
displayed mV’s to stabilize. Compare the readings to the Millivolt
Range Guidelines listed below.