t r o u b l e s h o o t i n g
g
No red test dye movement during SAMPLE FLOW test
•
If the red test dye is not moving through the pump tubing during
the
SAMPLE FLOW
test, inspect the electrode alignment in the
sensor module.
•
Confirm that the electrodes are installed in the correct order and
correctly seated.
•
Confirm that the compression plate is closed, and verify that the
bar at the rear of the sensor module is snapped inside the com-
pression plate hinge tabs. If not, snap the bar inside the hinge
tabs, then gently push down on the compression plate to engage
the spring around the fluid path exit port.
Release the compression plate.
•
Remove and inspect the pump tubing for poor connections, col-
lapsed walls or obstructions.
•
Replace the tubing when any of these conditions are evident, then
repeat the
SAMPLE FLOW
test with the red test dye (syringe
mode).