p r i n c i p l e s
o f
o p e r a t i o n
/
t h e o r y
r
Electrode Measurement/Diagrams
The PO2 electrode is electrically independent from the other electrodes
during measurements. The PCO2 electrode has an internal reference
electrode that is used during measurements, and is also independent
from the other electrodes.
Internal Fill
Solution
pH Sensitive Glass
Internal Reference
Electrode
Anode
Glass
Cathode
Internal Fill
Solution
Porous Spacer
Gas Permeable
Membrane
pH
PCO
2
PO
2
Na
+,
K
+,
Ca
++
, Cl-, pH
PCO
2
PO
2
The pH, Na
+
, K
+
, Ca++ and Cl- electrodes make all measurements
relative to the reference electrode. Errors in these electrodes may be
caused by problems in the electrode, reference electrode, or both.
The pH, PCO
2
, PO
2
, Na+, K+, Ca++ and Cl- electrodes used in the
EasyStat analyzer have no replaceable parts and are designed to
require no maintenance.