SAMPLER SWITCHES
Select
SAMPLER SWITCHES
to verify the sampler positioning
sensor and the sample mode switch.
CAPILLARY MODE?
is displayed. Lift the sampler to the capil-
lary mode and press YES.
SAMPLER CLOSED?
is displayed. Push the sampler down into
the closed position.
SYRINGE MODE?
is displayed. Lift the sampler to the syringe
mode and press YES.
RETURN SAMPLER
is displayed. Push the sampler down
into the closed position.
POSITION PASS
or
FAIL
and
MODE PASS
or
FAIL
are displayed. When the
SAMPLER
SWITCHES
test fails, refer to SENSOR MODULE in Troubleshooting.