Probe Wiper Replacement
Probe wipers should be changed every 3 months. In some cases they
may last longer, as the frequency of probe wiper replacement depends
on the number and type of samples analyzed. If the motion of the sam-
pler becomes difficult, wipe the probe with an oil wipe (provided in the
probe wiper replacement kit). If the sampler’s motion continues to be dif-
ficult, the probe wiper may need replacement.
Raise the sampler to the capillary sample position.
Using the tool supplied in the probe wiper replacement kit, turn the
probe wiper 90° to unlock. This will align the notches on the probe
wiper with the notch on the sampler.
Use the tool to pry the used probe wiper from the sampler.
Push the sampler down into the closed position, then raise the sampler
to the sample container mode. Wipe the probe with an oil wipe (pro-
vided in the probe wiper replacement kit).
Insert the new probe wiper with the notches aligned. With the tool, turn
the probe wiper 90° into the locked position. Lower and raise the probe
into the sample container mode several times to verify proper probe
motion.
0
1
4
7
2
5
8
3
6
9
yes
no
0
1
4
7
2
5
8
3
6
9
yes
no
Содержание EasyStat
Страница 1: ...EDIC EasyStat Operator s Manual...
Страница 5: ...Operator s Manual EDIC EasyStat Ref 7415 003770 901R1 2015 07...
Страница 10: ...4...
Страница 14: ...8 Front View 2 Analyzer Layout...
Страница 152: ...146...
Страница 154: ...148...
Страница 156: ......
Страница 157: ......