22
Calibrate
3
LIFT SAMPLER TO
ANALYZE
2
2
Hct Na
+
Ca
++
CALIBRATIO
AA
N
HOME MENU
1
*
2
*
3 CALIBRATE
4 DAILY CLEANER
5 SECOND MENU
HOME MENU
1
ANALYZE SAMPLE
2
ANALYZE QC
3 CALIBRATE
4 DAILY CLEANER
5 SECOND MENU
SLOPES
pH
61.02
PCO
2
53.17
PO
2
3.01
Hct
18.88
Na
+
58.98
K
+
61.50
Ca
++
24.62
CALIBRATION
BUBBLE DETECTOR CAL
air fluid
BD PASS 171 030
PUMP CAL
PUMP PASS 0799
REAGENT MODULE
STATUS
021%, 15 DAYS
JUNE-05-20
;13:35
4. Operating the Analyzer
s
The EasyStat analyzer requires a two-point calibration after instal-
lation and warm-up. Automatic two-point calibrations occur at
pre-selected intervals chosen under
CAL FREQUENCY
in the
CONFIGURATION
section of the
SETUP MENU
. Automatic
Hct calibrations are included in each two-point calibration. On-
demand two-point calibrations can be initiated in the
HOME MENU
by choosing
CALIBRATE
. If not calibrated, the
HOME MENU
only displays options 3, 4 and 5.
The ACCESS DOOR must be closed during warm-up, calibra-
tion, and sampling.
Select
CALIBRATE
to perform a two-point calibration and prepare
the EasyStat analyzer for sample analysis. The analyzer displays the
calibration status. Bubble detector and pump cal results are printed
at completion. After a successful calibration, the analyzer displays
and prints the slope of each electrode, the % of calibrants remaining,
and the number of days before expiration. After completion of a suc-
cessful calibration, the analyzer displays
LIFT SAMPLER TO
ANALYZE pH PCO2 PO2 Hct Na+ K+ Ca++
.
Ca++ is replaced with Cl- when Cl- mode is selected.