44
o p e r a t i n g
t h e
a n a l y z e r
2
SECOND MENU
1 STANDBY
2 REPLACE COMPONENTS
3 DIAGNOSTICS
4 STORED INFORMATION
5 SETUP MENU
6 ADJUST DISPLAY
REPLACE COMPONENTS
1 REAGENT MODULE
2 ELECTRODES
3 PUMP TUBING
4 SENSOR MODULE/SAMPLER
5 VALVE MODULE
no
no
Cl- MODE?
6
6 SWITCH TO Ca++/Cl-
PATIENT RESULTS
WILL BE LOST
CONFIRM?
yes
yes
RESTART THE ANALYZER
.
s
Replace Components
The EasyStat analyzer contains components which require periodic
replacement. These components include the electrodes (pH, PCO
2
,
PO
2
, Na+, K+, Ca++ , Cl-, and Reference) and reagent module. The
replacement schedule below must be followed to properly maintain the
EasyStat analyzer.
It is recommended that quality control material be analyzed after
component replacement to verify performance.
Switch to Ca++/Cl-, select
SWITCH TO Ca++/Cl-
. The
display will show either
Cl- MODE
(if the analyzer is currently in
Ca++) or
Ca++ MODE
(if the analyzer is currently in Cl- mode).
Press YES to move to next screen. Then press YES to confirm.
PATI
ENT RESULTS WILL BE LOST
. All the stored information
will be erased. The analyzer will restart in the selected mode. Replace
the correct electrode for the selected mode. Follow the instructions in
this chapter to operate the analyzer.