SLOPES
View the slope and calibration status for each electrode (
PASS
,
FAIL
or
INVALID
). Anelectrode with a
PASS
status indicates
the last calibration was successful and the electrode is currently cali-
brated. A
FAIL
status indicates that an error occurred during the
last calibration.
INVALID
indicates that the calibration status has
become invalid after the calibration; for example, the reagent module
has expired.
The acceptable slope ranges are:
pH
55.00 - 65.00
PCO2
46.50 - 60.00
PO2
1.00 - 5.00
Hct
13.00 - 22.00
Na+
52.00 - 64.00
K+
52.00 - 64.00
Ca++
20.00 - 34.00
Cl-
43.00 - 59.00