63
o p e r a t i n g
t h e
a n a l y z e r
r
s
Fuses
The EasyStat analyzer uses 250VAC fuses. When the voltage setting
is 110V, 0.8 amp fuses must be installed. When set to 220V, 0.4 amp
fuses must be installed. Correct fuses must be used with each volt-
age setting. To replace the fuse(s), remove the power cord, then push
down on the fuse holder release tab and pull straight out.
The EasyStat analyzer will only accept slow blow or time delay fuses.
Remove the fuse(s) from the holder and insert the new fuse(s). Confirm
that the fuse rating matches the voltage setting on the analyzer.
Reinstall the fuse holder, pushing in until it snaps into place. Connect
the power cord.
Contact your EasyStat dealer when repeated fuse replacement is
required.
H