t r o u b l e s h o o t i n g
g
SENSOR MODULE
WARMING UP
SENSOR MODULE
WARMING UP
SENSOR MODULE
SENSOR TEMPERATURE
This error message appears when the sensor module
has not stabilized at 37oC.
•
Make sure the door is closed. The heating circuits will not
function with the door open.
•
Select
SENSOR STATUS
from the
DIAGNOSTICS
menu.
•
Select
TEMPERATURE
to observe the real time sensor
and valve temperatures.
•
Close the access door and wait up to 30 minutes for
WARMING UP
to disappear.
•
If
WARMING UP
remains displayed after 30 minutes,
remove the sensor module, check and clean the contacts,
reinstall and wait up to 30 minutes.
•
If
WARMING UP
remains displayed after 30 minutes,
install a new sensor module, and wait up to 30 minutes.
•
If the temperature remains out of range, contact your
EasyStat dealer.
This error indicates an abnormally high temperature in
the sensor module.
•
Open the access door and carefully touch the sensor
module. Note if it feels unusually warm or hot. Wait a few
minutes for this message to disappear.
•
Close the access door.
•
If the
SENSOR TEMPERATURE
display remains,
power the analyzer off and then on.