95
s a m p l e
h a n d l i n g
a n d
c o l l e c t i o n
r
H
H
Evacuated tubes are not recommended for blood gas specimen collec-
tion due to variations in technique.
Capillary Sample
Capillary tubes are available from your EasyStat dealer. Follow
directions in capillary kit.
Errors occur from improper collection or handling procedures. When test
results are inconsistent with a patient’s present condition or test history,
consider the following sources of error: non-heparinized samples, incor-
rect heparin concentration, syringe filling speed, syringe mixing method,
sample storage technique, delayed sample analysis, multiple samples
from the same syringe, bubbles in the sample.
Interfering Substances
To test interferences, serum or whole blood was spiked with a poten-
tially interfering substance to the test concentrations shown in the
product specification. The interference was calculated using the differ-
ence between the medians of the spiked and unspiked samples.
The results found that the potassium thiocyanate 20 mg/dL affected
the chloride performance and caused chloride values increasing.