Used reagent module contains
biohazardous waste.
Reagent Module
If the EasyStat displays
RGNT MODULE EMPTY
or
RGNT
MODULE EXPIRED
, it is necessary to install a new reagent
module. Select
REAGENT MODULE
from the
REPLACE
COMPONENTS
menu. Fluid is automatically purged from the
sample path. The display prompts you to
REMOVE REAGENT
MODULE
. Open the access door. Push in the reagent module release
lever while holding the reagent module on the left side. Pull the mod-
ule to the left. When the guide arrow points to the edge of the reagent
module, pull the module straight out from the front of the analyzer.