Technical specifications
Electromagnetic compatibility (EMC)
54
GA 5752 2300 GB 26
8
8.7.2
Resistance to electromagnetic interference
Measurement /
Standard
Test level
Compliance level
Electromagnetic Envi-
ronment / Guidelines
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV Contact
discharge
± 8 kV Air discharge
± 6 kV Contact
discharge
± 8 kV Air discharge
Floors should be made of
wood or concrete or cov-
ered with ceramic tiles. If
the floor is covered with
synthetic material, then
the relative humidity has
to be at least 30 %.
Electrical fast tran-
sient disturbances /
burst
IEC 61000-4-4
± 2 kV for mains
cables
± 1 kV for input and
output cables
± 2 kV for mains
cables
± 1 kV for input and
output cables
The quality of the mains
power should correspond
to that of a typical com-
mercial or hospital envi-
ronment.
Surges (Surges)
IEC 61000-4-5
± 1 kV Differential
mode voltage
± 2 kV Common mode
voltage
± 1 kV Differential
mode voltage
± 2 kV Common mode
voltage
The quality of the mains
power should correspond
to that of a typical com-
mercial or hospital envi-
ronment.
Voltage dips, brief
interruptions and
fluctuations in the
supply voltage
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
) for
½ cycle
40 % U
T
(60 % dip in U
T
) for 5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
< 5 % U
T
(>95 % dip in U
T
) for 5
seconds
< 5 % U
T
(> 95 % dip in U
T
) for
½ cycle
40 % U
T
(60 % dip in U
T
) for 5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
< 5 % U
T
(>95 % dip in U
T
) for 5
seconds
The quality of the mains
power should correspond
to that of a typical com-
mercial or hospital envi-
ronment. If the user of
the product requires con-
tinued operation despite
interruptions in the power
supply, then it is recom-
mended that the unit be
operated with an uninter-
ruptible power supply or
battery.
Magnetic field of
the supply frequen-
cy (50 / 60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic fields gen-
erated at line frequency
should correspond to val-
ues typically found in
commercial or hospital
environments.
Note: U
T
is the AC mains voltage prior to application of test levels.
Tab. 9:
Resistance to electromagnetic interference