Maintenance
Malfunctions and troubleshooting
48
GA 5752 2300 GB 26
7
No. Malfunction
Cause
Remedy
3
Aspirator does not start
operation, operating status
display is not illuminated.
Equipment or mains plug is
not seated properly in the
socket.
Check the equipment and
mains plugs for proper
contact.
None or improper voltage.
Check the building fuse,
check the specifications on
the type plate.
The mains fuse is defective.
Netzsicherung
austauschen. [
Page 49]
4
Equipment cannot be
switched on and off.
The electronics are defective. Have the equipment
repaired by a service
technician authorised by
ATMOS.
5
Aspirator operates but
operating status display is
not illuminated.
LED on the operating status
display is defective.
Have the equipment
repaired by a service
technician authorised by
ATMOS.
6
Vacuum cannot be regulated. The membrane regulator is
defective.
Have the equipment
repaired by a service
technician authorised by
ATMOS.
7
Aspirator working but
vacuum gauge indicates no
vacuum.
The vacuum gauge is
defective.
Have the equipment
repaired by a service
technician authorised by
ATMOS.
8
Reduced / no flow rate
Septic fluid jar cap is not in
the correct position.
Position septic fluid jar cap
properly.
Hydrophobic filter is clogged
(vacuum gauge indicates
vacuum).
Replace hydrophobic filter.
Crack in the tube
Replace tube.
Seal is contaminated.
Replace seal.
Porous seal of septic fluid jar
cap
Replace seal.
Bent clamp, septic fluid jar
cap does not close.
Replace septic fluid jar
cap.
Septic fluid jar is full,
mechanical overflow
protection is closed (vacuum
gauge indicates vacuum).
Empty the septic fluid jar
and clean or replace septic
fluid jar and mechanical
overflow protection.
Mechanical overflow
protection is contaminated
with exudate.
Clean overflow protection
device or replace septic
fluid jar cap.
Tube connection in septic
fluid jar cap is clogged.
Clean the tube connection.