Cleaning and disinfection
Disinfection
46
GA 5752 2300 GB 26
6
Components
In solution
1
Wipe, spray disinfection
2
Aspirator basic equipment
X
Trolley
Catheter holder
Mains cable
Foot switch
Cap to cover the
equipment mount interface
Only spray disinfection
Bacterial filter cap
X
X
Float / float cage
Suction tube / connection
tube
Tube adapter
X
Housing of overflow
protection device,
universal / surgical
X
X
Hydrophobic bacterial and
viral filter (REF 5752 4514,
REF 5752 1783)
3
Disposable
Bacterial filter paper
When in use, must be replaced daily.
1 After exposure (as prescribed in the manufacturer's instructions), rinse components
thoroughly with water and dry them afterwards.
2 After exposure (as prescribed in the manufacturer's instructions) remove disinfectant residues
from the components using a moist cloth and dry them afterwards.
3 Replace the hydrophobic bacterial and viral filter immediately if it is discoloured, contaminated
or oversucked. Furthermore, the filter must be exchanged if the vacuum displayed is above
−0.3 bar / 30 kPa when the vacuum controller is in the ‘max’ position and the suction tube is
open.
Tab. 5:
Disinfection procedures