![MEDAP 5752 4856 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/medap/5752-4856/5752-4856_operating-instructions-manual_1759333043.webp)
Cleaning and disinfection
Cleaning
GA 5752 2300 GB 26
43
6
CAUTION!
Improper cleaning and disinfection can cause property damage!
Use only as much detergent and disinfectant as required.
CAUTION!
Improper cleaning and disinfection can cause property damage!
Perform visual and functional inspections after each cleaning and disinfection
process.
CAUTION!
Property damage due to changes in materials!
Almost all the components in the product are made of plastic. Solvents, some
disinfectants and some cleaning agents can soften plastic or cause tension
fissures.
Do not use alcohol-containing agents to clean the surfaces. Follow the instructions
for using disinfectants.
CAUTION!
Property damage due to tension cracks!
Acids or bases may cause tension cracks.
Do not treat polysulphone containers with strong acids or alkaline solutions.
NOTE
To clean the operating foil, rotate the regulation button to the left and unscrew to
remove. After cleaning, screw the regulation button back into place and rotate fully
to the right.
NOTE
Refer to the respective manufacturer's instructions for information on cleaning and
disinfection.
6.2
Cleaning
6.2.1
General
NOTE
Use only all-purpose cleaners which are slightly alkaline (soap solution) and which
contain surfactants and phosphates as the active cleaning agents.
In the event of heavily contaminated surfaces, use concentrated multi-purpose
detergent.
CAUTION!
Improper cleaning can cause property damage!
Residues of physiological saline solutions (e.g. sodium chloride) can attack the
surfaces of the product.
Remove residues of physiological saline solutions with a cloth dipped in clean
water. Then dry the product with a dry, lint-free cloth.