Operation
Suction
36
GA 5752 2300 GB 26
4
11
22
Fig. 25: Setting the vacuum level
Setting the vacuum level
Kink the suction tube on the patient side
and hold it closed. Set the vacuum with the
regulating knob (1) and check.
Increasing vacuum
Turn the regulator knob (1) to the right.
Read the value on the vacuum gauge (2).
Reducing vacuum
Turn the regulator knob (1) to the left.
Read the value on the vacuum gauge (2).
4.2.3
Operating the foot switch (only for BORA UP 2080 OP)
11
Fig. 26: Foot switch BORA UP 2080 OP
Operate the foot switch.
Aspirator is set to stand-by mode.
Yellow light emitting diode (1) is
illuminated.
Fußschalter erneut betätigen.
Yellow light emitting diode goes out.
Aspirator is set to operating mode.