Maintenance
General
GA 5752 2300 GB 26
47
7
7
Maintenance
7.1
General
Maintenance, repairs and period tests may only be carried out by persons who have the
appropriate technical knowledge and are familiar with the product. To carry out these measures,
the person must have the necessary test devices and original spare parts.
ATMOS recommends: Work should be carried out by an authorised ATMOS service partner. This
ensures that repairs and testing are carried out professionally, original spare parts are used and
warranty claims remain unaffected.
WARNING!
Health hazard!
The aspirator is used in the treatment of patients. The aspirator or parts of the unit
may be contaminated.
Prior to returning the aspirator for inspection or repair, remove the bacterial and
viral filter and all tubes and clean and disinfect the equipment.
7.2
Period tests
At least every 12 months a repeat test of the electrical safety should be performed according to
IEC 62353.
ATMOS recommends an inspection according to the manufacturer‘s specifications.
NOTE
After running for 3000 operating hours a check should be carried out on the unit.
This check must only be performed by authorised personnel.
7.3
Malfunctions and troubleshooting
No. Malfunction
Cause
Remedy
1
Aspirator does not start
operation, operating status
display is illuminated.
Vacuum is still present.
Switch off the aspirator,
turn regulating knob to the
left, switch on the
aspirator.
The motor is defective.
Have the equipment
repaired by a service
technician authorised by
ATMOS.
2
Absaugpumpe mit
Fußschalter läuft nicht an,
gelbe Leuchtdiode leuchtet.
Aspirator is in standby mode. Switch off standby mode
using the foot switch. The
aspirator starts operating.
If the foot switch is
operated again, the
aspirator switches back
into standby mode.